Validation of multi-frequency bioelectrical impedance analysis versus dual-energy X-ray absorptiometry to measure body fat percentage in overweight/obese Colombian adults

Robinson Ramírez-Vélez, Alejandra Tordecilla-Sanders, Jorge Enrique Correa-Bautista, Katherine González-Ruíz, Emilio González-Jiménez, Hector Reynaldo Triana-Reina, Antonio García-Hermoso, Jacqueline Schmidt-RioValle

Research output: Contribution to journalResearch Articlepeer-review

21 Scopus citations

Abstract

Objectives: To verify the validity of multi-frequency bioelectrical impedance analysis (mBCA) for predicting body fat percentage (BF%) in overweight/obese adults using dual-energy X-ray absorptiometry (DXA) as the reference method. Methods: Forty-eight adults participated (54% women, mean age = 41.0 ± 7.3 years old). The Pearson's correlation coefficient was used to evaluate the correlation between BIA and BF% assessed by DXA. The concordance between BF% measured by both methods was obtained with Lin's concordance correlation coefficient and Bland–Altman difference plots. Results: Measures of BF% were estimated as 39.0 (SD = 6.1) and 38.3 (SD = 6.5) using DXA and mBCA, respectively. The Pearson's correlation coefficient reflected a strong correlation (r =.921, P =.001). The paired t-test showed a significant mean difference between these methods for obese men BF% of −0.6 [(SD 1.95; 95% CI = −4.0 to 3.0), P =.037]. Overall, the bias of the mBCA was −0.6 [(SD 2.2; 95% CI = −5.0 to 3.7), P =.041], which indicated that the mBCA method significantly underestimated BF% in comparison to the reference method. Finally, in both genders, Lin's concordance correlation coefficient showed a strong agreement. More specifically the DXA value was ρc = 0.943 (95% CI = 0.775 to 0.950) and the mBCA value was ρc = 0.948 (95% CI = 0.778 to 0.978). Conclusions: Our analysis showed a strong agreement between the two methods as reflected in the range of BF%. These results show that mBCA and DXA are comparable methods for measuring body composition with higher body fat percentages. However, due to broad limits of agreement, we can only recommend mBCA for groups of populations.

Translated title of the contributionValidación de análisis de impedancia bioeléctrica multifrecuencia versus absorciometría de rayos X de doble energía para medir el porcentaje de grasa corporal en adultos colombianos con sobrepeso/obesidad.
Original languageEnglish (US)
Article numbere23071
JournalAmerican Journal of Human Biology
Volume30
Issue number1
DOIs
StatePublished - Jan 1 2018